3–7 Jul 2011
ETH Zurich, Switzerland
Europe/Zurich timezone

EIGER a new single photon counting detector for X Ray applications: performance of the chip.

5 Jul 2011, 12:05
20m
HG E7 (ETH Zurich, Switzerland)

HG E7

ETH Zurich, Switzerland

www.ethz.ch www.psi.ch
Oral presentation Detector Systems Detector Systems I

Speaker

Valeria Radicci (Paul Scherrer Institut)

Description

EIGER is the next generation of single photon counting detector for synchrotron radiation designed by the PSI-SLS detector group. It features a pixel size of 75x75 μm2 and frame rates up to 22kHz. An array of 256x256 pixels fits on a 2cm x 2cm chip. The chip provides 4, 8 and 12 bit counting modes and practically an infinite dynamic range (32 bits) due to the continuous read/write and the summation of frames on the fly in firmware. Along with the first X Ray absorption images, the characterization and performance of the chip will be presented. The most important parameters i.e. energy calibration, noise, minimum energy threshold and rate capability were measured on a single chip system with an X Ray tube and at the SLS-PSI synchrotron. Trimming and irradiation studies will be discussed. Tests so far have shown that the EIGER system meets its specifications. A report on the status and plan for a full module (2x4 chips) and 4M pixel EIGER detector will also be given.

Primary author

Valeria Radicci (Paul Scherrer Institut)

Presentation materials