Skip to main content

Quartet Meeting in Paris

Europe/Zurich
room 210 (Laboratoire Kastler Brossel)

room 210

Laboratoire Kastler Brossel

Jussieu campus, hallway 13-23, meeting room 210 next to the LKB coffee room.
  • Monday 7 November
    • 10:00 10:10
      Welcome and Overview

      Reminder of QUARTET goals, timeline, and goals for meeting

      Convener: Nancy Paul (Laboratoire Kastler Brossel)
    • 10:10 11:00
      Report on recent MuX results and status of PSI beamline 50m
      Speakers: Frederik Wauters (Johannes Gutenberg University Mainz), Katharina von Schoeler (PSI - Paul Scherrer Institut)
    • 11:00 11:55
      Moments of the charge distribution 55m
      Speaker: Ben Ohayon (ETH Zurich)
    • 12:10 13:40
      Lunch 1h 30m
    • 13:55 14:55
      How to measure?: Needs for MMC before beamtime--calibration, testing

      Plan experimental setup for 2023

      Convener: Andreas Fleischmann
      1. The signal size is a function of the bath temperature. They record the temperature for each event and correct for the fluctuations.
         
      2. Some events make a peak to the left of the main line. Could be improved.
         
      3. The rise time would be around a microsecond in our application
         
      4. Order of few ms of decay time is a good compromise for low rate experiments like ours.
         
      5. The signal analysis is not so sensitive to the filtering mechanism.
         
      6. For the MaXs-20, can see up to 60keV but the resolution deteriorates above 10keV.
         
      7. For the larger detectors, the response is more linear.
         
      8. When averaging out pixel, it is typically ok to sum up scaled pixels.
         
      9. MaXs-100, is 10x10, around 40eV resolution. can go up to around 200keV.
         
      10. Takes about 2.5 days between disassembling at Heidelberg and pressing "start" at GSI.
         
      11. Some calibration linecenters depend on the chemistry. They like gamma lines, which some to agree to around 0.5eV.
         
      12. The deviation from better calibration are ascribed to effects of ADC nonlinearity. 
         
      13. There may be better ADC around. 
         
      14. "in principle you can eliminate this problem on a voltage level".
         
      15. This limits the ability to do "sparse calibration". We probably need close calibration lines until ADC upgrade.

       

      Setup:
       

      1.  

       

       

    • 14:55 15:25
      Lessons learned with TES detectors at J-PARC 30m
      Speaker: Nancy Paul (Laboratoire Kastler Brossel)
    • 15:25 16:25
      How to measure?: Experimental setup at PSI

      Plan experimental setup for 2023

      Convener: Andreas Knecht (Paul Scherrer Institut)
      1. The signal size is a function of the bath temperature. They record the temperature for each event and correct for the fluctuations.
         
      2. Some events make a peak to the left of the main line. Could be improved.
         
      3. The rise time would be around a microsecond in our application
         
      4. Order of few ms of decay time is a good compromise for low rate experiments like ours.
         
      5. The signal analysis is not so sensitive to the filtering mechanism.
         
      6. For the MaXs-20, can see up to 60keV but the resolution deteriorates above 10keV.
         
      7. For the larger detectors, the response is more linear.
         
      8. When averaging out pixel, it is typically ok to sum up scaled pixels.
         
      9. MaXs-100, is 10x10, around 40eV resolution. can go up to around 200keV.
         
      10. Takes about 2.5 days between disassembling at Heidelberg and pressing "start" at GSI.
         
      11. Some calibration linecenters depend on the chemistry. They like gamma lines, which some to agree to around 0.5eV.
         
      12. The deviation from better calibration are ascribed to effects of ADC nonlinearity. 
         
      13. There may be better ADC around. 
         
      14. "in principle you can eliminate this problem on a voltage level".
         
      15. This limits the ability to do "sparse calibration". We probably need close calibration lines until ADC upgrade.

       

      Setup:
       

      1.  

       

       

    • 16:25 16:45
      Coffee 20m
    • 16:45 17:15
      What will we measure?: Theoretical contributions to systematics and uncertainties
      • 16:45
        Discussion on letter of interest 30m
        Speaker: Andreas Knecht (Paul Scherrer Institut)
    • 17:15 17:45
      What will we measure?: Experimental contributions to systematics and uncertainties
    • 17:55 18:55
      Discussion
    • 19:00 21:30
      Drinks and Dinner--LE PETIT PONTOISE 2h 30m

      9 Rue de Pontoise
      75005 Paris, France

      http://www.lepetitpontoise.fr/

  • Tuesday 8 November
    • 10:00 10:30
      Funding for 2023 and beyond 30m
      Speakers: Ben Ohayon (ETH Zurich), Frederik Wauters (Johannes Gutenberg University Mainz), Nancy Paul (Laboratoire Kastler Brossel)
    • 10:30 11:00
      Structure of Test beam time request
      Convener: Andreas Knecht (Paul Scherrer Institut)
    • 11:00 12:55
      Planning and to-do list 1h 55m

      Distribution of tasks for LOI to request test beam in 2023

      Speaker: Nancy Paul (Laboratoire Kastler Brossel)